188W/188Re Generator
Not for Human Use.
Elution
The elution yield of rhenium-188 is 75-85 %, and it remains stable throughout the shelf-life of the generator.
Set up
- The 188W/188Re generator must be eluted with sterile 0,9 % saline solution only.
- The maximum time for the 188W/188Re generator to be left uneluted is three (3) weeks.
Typical 188W/188Re generator elution profile
Specifications
Column material | Aluminium oxide (Al2O3) packed in quartz glass column |
Eluent | Sterile 0.9% saline solution |
Chemical form | Sodium perrhenate (Na188ReO4) |
Appearance | Clear and colorless solution |
Elution volume | 5-14 ml (with small volumes not completed elution) |
Elution yield | 75-85% |
Radiochemical purity | >98% Re(VII) |
Tungsten breakthrough | 188W <0.01% in relation to 188Re at calibration time |
Total gamma impurities | <0.01% in relation to 188Re at calibration time |
Shelf-life | 10 months |
Specific activity | Product is carrier free |
Generator size (nominal 188W activity) | 3.7-150 GBq |
Generator dimensions Diameter | 145 mm, height 210 mm |
Generator weight | 16 kg |
Interface Standard | Stainless steel needles for elution with evacuated vials. On request: Tubings with Luer-Lock connectors (inlet female, outlet male) |